Skip to main content

Table 2 Between day reliability of peak torque, work done and endurance ratio in all patients and volunteers (n = 39).

From: Use of a Cybex NORM dynamometer to assess muscle function in patients with thoracic cancer

 

Mean value for the tests

Mean difference between the tests

SD of the difference between tests

Intraclass correlation coefficient (95% CI)

Quadriceps

    

Peak torque (N)

69

9

14

0.91 (0.82–0.95)

Work (J)

1098

125

321

0.84 (0.69–0.92)

Endurance ratio (%)

87

-15

38

0.56 (0.17–0.77)

Hamstrings

    

Peak torque (N)

42

8

13

0.86 (0.75–0.93)

Work (J)

682

110

197

0.88 (0.77–0.94)

Endurance ratio (%)

102

-14

60

0.49 (0.02–0.74)

  1. SD = standard deviation; CI = confidence interval